Brand: FLUKE BIOMEDICAL

Digital Electrical Safety Tester Fluke Biomedical LT544

Code: Fluke Biomedical LT544


 

Description

Features LT544Dlite Digital Safety Tester

  • Small, lightweight, self-contained portable instrument
  • Universal Power Supply. Input voltage from 85 - 265 VAC
  • Chassis Resistance measurements are made with a four-wire Kelvin bridge to eliminate errors due to cable length
  • Test method complies with AAMI load per ANSI/AAMI ES1-1993
  • Leakage Current can be measured in Open/Closed Neutral, Normal/Reversed Polarity and Open/Closed Ground
  • All Leakage Current measurements are displayed directly in µAmps

 

 

Features LT544Dplus Digital Safety Tester

  • Small, lightweight, self-contained portable instrument
  • Universal Power Supply. Input voltage from 85 - 265 VAC
  • Chassis Resistance measurements are made with a four-wire Kelvin bridge to eliminate errors due to cable length
  • Test method complies with AAMI load per ANSI/AAMI ES1-1993
  • Leakage Current can be measured in Open/Closed Neutral, Normal/Reversed Polarity and Open/Closed Ground.
  • All Leakage Current measurements are displayed directly in µAmps.
 

 

Specifications LT544Dlite Digital Safety Tester

Ground Resistance

Range:

 0 - 19.99 ohms

Accuracy:

 ± 1% R, ± 1 LSD

Resolution:

 10 mOhms

Leakage Current

Range:

 0 - 1, 999 µA

Accuracy:

 DC & 25 Hz - 1kHz: ±1% of reading, ± 3LSD

 

 1.0KHz - 100 KHz: ±2.5% of reading, ± 3LSD

 

 100 KHz - 1 MHz: ±5% of reading, ± 3LSD

Current Capacity

LT544DLite- 1515

 15Amps at 120 VAC

LT544DLite- 230V

 10Amps at 230 VAC

LT544DLite- 1520/2020

 20Amps, at 230 VAC, 20% duty cycle (2 minutes on, 8 minutes off)

Controls

 

 1. Function Switch - 4 position rotary

 

 2. Neutral Switch - 2 position rocker (Open/Closed)

 

 3. Polarity Switch - 3 position rocker (Normal/Off/Reversed)

 

 4. Leakage Switch - 2 position rocker (Chassis/Earth)

DUT Power

 

 85 - 265 VAC, 16 Amps@ 120 V, 10 Amps @ 230 V

Physical Size

 

 3.5 x 6.2 x 1.6 inches (90 x 180 x 38 mm)

 

 

 

Specifications LT544Dplus Digital Safety Tester

Line Voltage

Range:

 85 V ac to 265 V ac

Resolution:

 1 Volt

Accuracy:

 2 % Reading ± 1 LSD

Instrument Current

Range:

 0 Amps to 19.99 Amps

Resolution:

 10 mA

Accuracy:

 5 % Reading ± 1 LSD

Ground Resistance

Range:

 0 ohms to 19.99 ohms

Accuracy:

 ± 1 % Reading, ± 1 LSD

Resolution:

 10 mOhms

Load Current

Range:

 0 to 14.99A (LT544Dplus - 1515)

 

 0 to 19.99A (LT544Dplus-2020)

 

 0 to 9.99A (LT544Dplus-230V)

Accuracy:

 ± 4 % Reading, ±1 ISD

Leakage Current

Range:

 0 µA to 1, 999 µA

Accuracy:

 DC&25 Hz to 1 kHz: ± 1 % Reading, ± 3 LSD

 

 1 KHz to 100 KHz: ± 2.5 % of Reading, ± 3 LSD

 

 100 KHz to 1 MHz: ± 5 % of Reading, ± 3 LSD

Current Capacity

 

 LT544Dplus-1515: 15 Amps at 120 V ac

 

 LT544Dplus-230V: 10 Amps at 230 V ac

 

 LT544Dplus-2020: 20 Amps, at 120 V ac, 20 % duty cycle (2 minutes on, 8 minutes off)

Controls

 

 1. Function Switch - 4 position rotary

 

 2. Neutral Switch - 2 position rocker (Open/Closed)

 

 3. Polarity Switch - 3 position rocker(Normal/Off/Reversed)

 

 4. Leakage Switch - 2 position rocker (Chassis/Earth)

DUT Power

 

 85 V ac to 265 V ac, 16 Amps @ 120 V, 10 Amps @ 230 V

Power Connectors

 

 LT544Dplus ....................Standard US

 

 LT544Dplus-EUR ..........Schuko

Physical Size

 

 90 mm x 180 mm x 38 mm (3.5 in x 6.2 in x 1.6 in)

 

More products FLUKE BIOMEDICAL

HANDHELD ELECTRICAL SAFETY ANALYZER FLUKE BIOMEDICAL 175

• Handheld portability (fitting in a briefcase)
• Self-switching 120/240 V operation
• Patient-lead-leakage testing
• 15 A and 20 A capabilities
• Digital accuracy
• Dual-lead testing

 

DIGITAL RADIATION MONITOR FLUKE BIOMEDICAL PRIMALERT 05-443/ 05/444

• LED digital display
• Low and High alarm indicators
• Programmable alarm indicator
• Optional remote
• Detector Fail indicator
• Battery backup
• Data output/RS-232

 

X-RAY RADIATION DOSIMETER TEST SET TOOLS FLUKE BIOMEDICAL TNT 12000

• Choice of all-in-one detector, dosimeter, integral ma/mAs and handheld display or user’s own laptop interface.
• 100 % ZigBee wireless operation
• Compact handheld design
• Sets up in seconds
• Displays all values in one shot (all-in-one detector)

 

ADVANCED DETECTION METER FOR ALPHA, BETA, GAMMA, X-RAY RADIATION FLUKE BIOMEDICAL ASM-990

• Auto-scaling measurement of rate and dose simultaneously, with the capability to record peak rate
• Up to [5] different probes can be calibrated with one unit
• Survey Mode data logging feature allows user to store up to 5 separate survey sequences
• Saved data can be uploaded to a PC via included Infrared Data (IrDA) transmitter
• Easy-to-use multifunction keypad for intuitive menu navigation
• Backlit analog/digital LCD display with full-range audio output capability

 

ULTRASOUND ELECTRICAL SAFETY TRANSDUCER LEAKAGE CURRENT TESTER FLUKE BIOMEDICAL ULT800

• Handheld
• Stand-alone operation
• Direct measurement of leakage current
• Battery-operated
• Independent of 120 V or 240 V systems
• Built-in self-test circuit
• Auto shut-off to conserve battery

 

OXIGEN O2 CONCENTRATION ANALYZER FLUKE BIOMEDICAL MAXO2 PLUS AE

• One-touch calibration, with reminder
• Long battery life (approx. 5,000 hrs)
• Impact resistant and drip proof
• External MAX-250E Oxygen Sensor

 

PHOTOTHERAPY RADIOMETER LIGHT RADIATION MEASURING FLUKE BIOMEDICAL DALE40

• Large LCD
• Accurate to ±5% of full scale
• Spectral range of 429-473 nm
• Complete portability


ELECTRICAL SAFETY ANALYZER FOR MEDICAL ROOMS FLUKE BIOMEDICAL ESA612

• Portable, ergonomic, lightweight, and easy to use
• Large, easy-to-read display with adjustable contrast
• Human-factors-designed user interface
• Tilt stand design for stand-up testing in field environments
• Five applied parts jacks and easy ECG snap connection

 

UNIVERSAL DIGITAL RADIATION MONITORING RATEMEMTER FLUKE BIOMEDICAL 942A/ 946A/ 952A

• User adjustable alarm set-points; retained indefinitely on power outage, keylock protection available
• Front panel pushbuttons for check source activation, high voltage indication, alarm setpoint indication, alarm acknowledge, and power ON/OFF
• Small front panel size allows dedicated display for each channel while minimizing panel space
• Ability to use existing cables when replacing most analog readouts

 

DEFIBRILLATOR PULSE EXTERNAL PACER ANALYZER FLUKE BIOMEDICAL 6000D/ 7000DP

• Lown, Edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility
• AED technology compatibility
• First-class measurement accuracy ± 1 % of reading + 0.1 J
• Intuitive user interface and backlight, easy-to-read display
• Portable, rugged, easy to carry