Magnetic flux
-
Magnetic flux density
-
Magnetic field strength
-
Magnetic moment
-
Magnetic potential
-
Polarization
-
Analogue voltage integrator for measuring all quality-related magnetic parameters
-
Integrated calibration device
-
Automatic probe identification
-
Selection of measuring units
-
Integrated comparator
-
Self-testing function
-
Peak value memory
-
Analogue output
-
Well-structured user interface
-
Illuminated graphic display
-
Demonstration software for PC connection
Measuring all magnetic properties of magnetic material relevant to quality control.
Measuring all magnetic properties of magnetic material relevant to quality control.
Applications in e.g. research and development, automatic process control, quality control, and at incoming inspection.
When the flux density is changed a voltage is induced proportional to the change in flux. The Fluxmeter F 10 integrates this voltage and indicates the voltage integral.
The self-calibrating function of the instrument ensures that the measurements are always accurate. Display of the measured values in a selection of scales: e.g. Tesla, A /m or Vs.
In this way optimal use is made of the possible applications.
User-friendly features:
- Drift compensation
- Digital display filter
- Conversion of measuring scales
- Self-calibration Window comparator
- Percentage display
- Physical probe parameters saved in memory
- Automatic probe identification
- Choice of manual or automatic range
- Remote control via digital interfaces
Technical data: |
|
Measuring mode: |
free of drift integrator with 24-bit A / D-conversion |
Measuring ranges: |
400 µVs, 4, 40, 400 mVs |
Resolution: |
maximum 0,1 µVs |
Selection of range: |
automatic or manual |
Accuracy: |
0,2 % after self-calibration |
Measuring rate: |
50 measurements / sec |
Internal calibration: |
voltage- / time reference |
Drift: |
<± 1 µV s / min |
Input resistance: |
1 kOhm, 10 kOhm ± 0,1 % |
Connections: |
thermal low voltage mini-sockets on the front and back panel |
Display: |
LCD 100 x 80mm, 4-digit display of measured value, bar graph display |
Digital output: |
low / high / in / run |
Digital input: |
reset / hold / range / drift zero |
Interfaces: |
RS-232 C, USB optional |
Analogue outputs:1x galvanically isolated, ± 10 V 1x configurable (average value, min. / max., peak / peak, ± 10 V) |
|
Peak value memory: |
digital recording at intervals of 20 ms, analogue recording for high-speed pulses (optional) |
Dimensions: |
120 x 250 x 250 mm (height x width x depth) |
Mains supply: |
105– 125 V AC/ 210 – 250 V AC, 50 / 60 Hz |