Three-Channel Precision Laboratory Teslameter Brockhaus GM-460 testing magnetic

    office@distek.ro     ROM     ENG  
Login | Create account | Forgot password?
Product cart  Products: 0 | Total: 0    
 
 
    Home |  About Us |  News |  Promotions |  Discounts |  Services |  Tech Library |  How to order? |  Sitemap |  Contact
 
 
Products         [+ Expand]   [- Colapse]

Electromagnetic Field Testers > BROCKHAUS (11) > Three-Channel Precision Laboratory Teslameter Brockhaus GM-460

 


  ..  Products navigator [5] from [11]    ..
 

Three-Channel Precision Laboratory Teslameter Brockhaus GM-460

Code: Brockhaus GM-460
 
BROCKHAUS
 

» Technical Specs

  • Three-channel Gaussmeter for determining the magnetic field strength of permanent magnets and systems
  • Measurement of magnetizing pulses
  • User-friendly measuring instrument
  • Mains supply
  • Application in laboratories and in production lines
  • Field and temperature measurement via the probe input
  • Automatic temperature compensation
  • Application as a single-, two- or three-channel Gaussmeter

 

Operating principle

Multi-channel Gaussmeter with a resolution of 5 3/4 digits. Up to a maximum of three probes can be connected simultaneously. Each measuring channel is shown in one line in the display. Measurement of Hall voltage created in the Hall sensor which is proportional to the adjacent magnetic field.

All the functions are displayed and controlled via the front panel of the instrument. Input via keyboard, readings on a 4-line fluorescent display with alphanumerical text. Measuring range, zero, and functions such as max hold, relative or alarm, can be set individually for each channel.

Two analogue outputs provide high accuracy and the possibility to analyse the development of the field. The corrected analogue output shows the exact value displayed, the non-corrected analogue output shows the field strength in real time.

The comprehensive Hall probe programme includes over 500 different measuring probes, as well as conventional transverse and axial probes and probes for bi-and tri-axial measurements. BROCKHAUS also supplies probes for low temperature applications and for measuring minimal magnetic fields.

 

Technical data:

Gaussmeter 460

Measuring mode:

measurement of Hall voltage

Measuring range:

300 mG, 3 G, 30 G, 300 G, 3 kG, 30 kG, 300 kG

Measuring frequency:

DC, 10 Hz - 400 Hz

Measuring accuracy:

± 0,1 % of DC value, ± 2 % of AC value
0,005 % of measuring range

Repeatability:

± 0,0007 %

Measuring rate:

4 measurements/sec / display
18 measurements/sec / interface

Selection of measuring range:

automatic or manual

Resolution:

5 3/4 digits (1 of ± 300.000)

Display:

4-line, vacuum-fluorescence

Probe type:

all laboratory probes

Temperature range:

0°C - 40°C

Analogue outputs:

corrected and non-corrected ± 10 V

Interfaces:

RS-232, IEEE-488

Power supply:

100 - 240 V AC, 50 / 60 Hz

Functions:

Max Hold, Filter, Alarm, Zero Probe, Relative Reading, Auto Range, Memory Hold, AC/DC, Gauss, Tesla, Keylock, Pulse Field Measurement

Accessories:

Zero gauss chamber, power cable, operating manual

Special accessories:

probe cable in lengths 3 / 7,6 / 15,24 / 30,49 m
305 mm zero chamber / interface cable
19" drawer



» Standard included accessories

  • Zero chamber,
  • power cable,
  • operating manual


» Recommended Options, Services and Accessories

 
probe cable in lengths 3 / 7,6 / 15,24 / 30,49 m
305 mm zero chamber / interface cable
19" drawer
 

Hall probes overview Gaussmeter

 

Axial Probes

Model

L

D

A

Stem Material

Corrected accuracy (%) of reading

Temperature coefficient (maximum) zero

Temperature coefficient (maximum) calibration

MMA-2502-VH

2 " ± 0.063"

0.25" ± 0.006''

0.010"

Aluminum

± 0.25 % to 30 kG

± 0.09 G/°C

- 0.04 %/°C

MNA-1904-VH

4' ± 0.125"

0.187" ± 0.005''

0.005''

Fiberglass
epoxy

MMA-2502-VG

2" ± 0.063''

0.25" ± 0.006''

0.010"

Aluminum

± 0.15 % to 30 kG

± 0.13 G/°C

- 0.005 %/°C

MNA-1904-VG

4' ± 0.125"

0.187" ± 0.005''

0.005"

Fiberglass
epoxy

 

Transverse Probes

Model

L

D

A

Stem Material

Frequency Range

Corrected accuracy (%) of reading

Temperature coefficient (maximum) zero

Temperature coefficient (maximum) calibration

MMT-6J04-VH

4"± 0.125"

0.061" max

0.180"

Aluminum

DC

± 0.25 % to 30 kG

± 0.09 G/°C

- 0.04 %/°C

MNT-4E04-VH

0.045" max

0.150''

Rigid glass epoxy

DC and 10 Hz to 400 Hz

MMT-6J04-VG

0.061" max

0.180"

Aluminum

DC

± 0.15 % to 30 kG

± 0.13 G/°C

- 0.005 %/°C

MNT-4E04-VG

0.045" max

0.180"

Rigid glass epoxy

DC and 10 Hz to 400 Hz

 

Flexible Transverse Probes

Model

W

T

Active Area

Stem Material

Frequency Range

Corrected accuracy (%) of reading

Temperature coefficient (maximum) zero

Temperature coefficient (maximum) calibration

MFT-3E03-VH

0.135" max

0.025" max

0.040" dia (approx)

Flexible
Tubing

DC and 10 Hz to 400 Hz

± 0.25 % to 30 kG

± 0.09 G/°C

- 0.04 %/°C

MFT-3E03-VG

± 0.15 % to 30 kG

± 0.13 G/°C

- 0.005 %/°C

 

 

Home |  About Us |  News |  Promotions |  Discounts |  Services |  Tech Library |  Terms and Conditions |  How to order? |  Useful links |  Sitemap |  Contact |  Politica Cookie

        