Entry level biomass analyser with a 5 year life high range CO sensor
Features
Applications
Measures
Calculates
Fuels Types
Box Contents
Specification
Parameter
Range
Resolution
Accuracy
Temp Measurment
Flue Temperature
0-600°C
0.1°C
±2°C±0.3% reading
Inlet Temperature(Internal Sensor)
0-50°C
±1°C±0.3% reading
Inlet Temperature(External Sensor)
Flue Gas Measurement
Oxygen*2
0.21%
0.1%
±0.3%
Carbon Monoxide(H2 compensated)
0-20ppm21-4000ppm nom4,001-10,000ppm20,000ppm max for 15 mins
1ppm
±3ppm±5% of reading±10% of reading
Carbon Dioxide*1
0-20%
±0.3% volume
Efficiency (Net or Gross)*2
0-99.9%
±1.0% reading
Efficiency High (C)*2
0-119.9%
Excess Air*2
0-250%
±0.2% reading
CO/CO2 ratio*2
0-0.999
0.0001
±5% reading
*1 Using dry gases at STP*2 Calculated
Pre-programme fuels
Natural gas, Light Oil, Propane, Butane, LPG, Wood Pellets.
Dimensions
Weight
1kg / 2.2lb
Handset
200mm / 7.9" x 45mm / 1.8" x 90mm / 3.5"
Probe
L300mm / 11.8" x Dia 6mm / 0.25" with 200mm / 7.8" long stainless steel shaft, type K thermocouple and 3m / 6ft long neoprene hose
Ambient Operating Range
±0°C to ±40°C / 32-104°F10% to 90% RH non-condensing
Storage Capacity
20 sets of test results in volatile memory
Battery Life
4AA cells> 8 hours using Alkaline AA cells
Charger (optional)
110Vac/220 Vac nominal for NMiH batteries only12V in vehicle charger for NMiH batteries only